To calibrate the oscilloscope

Equipment Required
Equipment
Cable (2)
Adapter
Adapter
Adapter

Set up the equipment

C A U T I O N
The effects of ELECTROSTATIC DISCHARGE can damage components.
Use grounded wrist straps and mats when you are performing any kind of
service on this module.
The HP 16533/34A oscilloscope module contains a calibration
PROTECT/UNPROTECT switch. This switch must be set to UNPROTECT
before new calibration values from the operational accuracy calibration can
be stored to nonvolatile RAM.
HP 16533/34A multi-card modules must be separated into single card
modules before performing the operational accuracy calibration on the
acquisition circuitry.
Remove power from the HP 16500 mainframe. If a multi-card module
1
is being calibrated, disconnect all Master/Expander Trigger Cables
from the ECL EXT TRIG IN and OUT connectors on the rear panel of
each HP 16533/34A card.
Pull out halfway all cards and filler panels located above the
2
oscilloscope, then remove the filler panels and cards located above the
oscilloscope.
11–10
Critical Specification
BNC, 9-inch (equal length)
BNC (m)(m)
BNC tee (m)(f)(f)
BNC (f)/SMB (m)
Recommended
Qty
Model/Part
HP 8120-1838
2
HP 1250-1236
1
HP 1250-0781
1
HP 1250-1236
1