Advanced Options for CMU Calibration
Calibration
This dialog box is opened by clicking the Advanced Options... button, and is used to set the
information required to measure the open/short/load correction data and the phase
compensation data. This dialog box provides the following action button.
OK: Applies the setup changes and closes this dialog box.
Cancel: Cancel the setup changes and closes this dialog box.
Frequency: You can select the measurement frequency setup mode from the following modes.
Measurement frequency setup modes:
Default (Use factory specified frequencies): Usually, select this mode. Twelve points are
automatically set. They are 1 k, 2k, 5 k, 10 k, 20 k, 50 k, 100 k, 200 k, 500 k, 1 M, 2 M, and 5 MHz.
Specifies by list: Select this mode when you want to set the frequencies independently. The
frequency must be 1 kHz to 5 MHz. The number of frequencies must be 1 to 101. Click Add to
open a dialog box, and enter the value. For the unnecessary value, highlight the value and click
Delete.
Specifies by range: Select this mode when you want to set the frequencies sequentially and
automatically. Specify the start frequency, the stop frequency, the number of steps, and the scale
LINEAR/LOG. The start frequency must be less than the Stop value and at least 1 kHz. The stop
frequency must be more than the Start value and no more than 5 MHz. The number of steps must
be 2 to 1001.
NOTE:
If the device measurement frequency is not equal to the correction data measurement frequency,
calculation will be performed automatically to get the correction data for the device measurement
frequency, and the calculated correction data will be used for the data correction of the
measurement data.