To Set Auto Analysis Function
Auto Analysis Setup
Click the Auto Analysis Setup tab to display the auto analysis setup screen. The auto analysis
capability allows you to display/move the marker/line on the data graph automatically after
measurement. The above example displays marker at Vd=1 V point and draw a tangent line for
this point.
Interpolation Mode: Check this box to enable the interpolation mode. You can position marker
between measurement points.
Line 1, Line 2: This area sets the line 1 or line 2 automatically displayed and moved to the
specified position after the measurement. Check Enable box to enable the line.
For Type, select a line type, Normal, Gradient, Tangent, or Regression.
For Axis, select the axis effective for the line.
In First Point and Second Point, specify the cursor points used to draw line.
In Gradient, specify the slope of the Gradient line.
Marker: This area sets the marker automatically displayed and moved to the specified position
after the measurement. Check Enable box to enable the marker.
For Condition, enter a variable name and a condition expression to specify where you want the
marker to display.
For After, check this box to enable the additional condition. And enter a second variable and
condition expression. This sets up a search start condition for finding specified point.
For example, you can specify the following expressions to search for a measurement point that
satisfies the first condition after the second condition is satisfied.
Condition: [DGM] = [MAX(DGM)*0.01] [x] After [DGM] = [MAX(DGM)]
Note:
The Apply button performs calculation of the definition in this screen. If this button is clicked
when the Data Display window associated with this test has been opened, the display is also
updated. The associated Data Display window will be the Data Display window that displays the
last measurement data or that is recalled at last.