To Set Meas. Timing and Abort Function
Measurement Setup
Timing parameters
At the Integration Time area, you set the integration time of the impedance measurement by the
MFCMU.
•Mode: AUTO or PLC. Rule to decide the integration time. For details, see online help or User's
Guide.
•Factor: Coefficient for the reference value of the integration time.
At the Timing area, you set the source output and measurement timing.
•Hold: Source output hold time. Time after the measurement trigger until starting delay time.
•Delay: Measurement start delay time. Time after the hold time until starting measurement.
The Sweep [ ] status field sets the sweep abort function.
When this field is CONTINUE AT ANY, the sweep measurement is not aborted.
When this field is STOP AT ANY ABNORMAL, the sweep measurement will be aborted when one
of the following conditions is detected.
•Measurement range overflow
•MFCMU error
•SMU compliance
•SMU oscillation
Integration time (AUTO or PLC)
5-40
Sweep abort function
CONTINUE AT ANY
STOP AT ANY ABNORMAL
-Range overflow
-MFCMU error
-SMU compliance
-SMU oscillation
Trigger
Measurement time
Delay time
Hold time