To Define Switch Setup
-Open All
-Switch Setup
-Close Range
-Open Output
-Open Range
-Open Input
This setup screen is used to create the relay control setup (switch setup) for Agilent
B2200A/B2201A Switching Matrix.
To send the switch setup to B2200, click the Single button. Before doing this, you have to
establish the GPIB connection to B2200 by using the Switching Matrix tab screen of the
Configuration window. See Module 8.
This screen provides the following GUI.
Add: Adds a setup row for relay connection.
Delete: Deletes the setup row. To select the setup row, use the left radio button.
Up: Moves the selected setup row upward.
Down: Moves the selected setup row downward. The setup row provides the following entry
fields.
Operation: Relay control operation. Open All, Switch Setup, Close Range, Open Output, Open
Range, or Open Input. See the next page.
Input: B2200 input port number or label. Available for the Switch Setup, Close Range, and Open
Input operations.
Outputs: B2200 output channel number or label. Four entry fields are available for the Switch
Setup and Open Output operations to specify the output channels. Two entry fields are available
for the Close Range and Open Range operations to specify the range of output channels.
Rule: Relay connection rule. BBM or MBBR.
•BBM: Breaks the previous connection and then makes the new connection for the specified
outputs.
•MBBR: Makes the new connection and then breaks the previous connection for the specified
outputs.
Setup Name
-BBM
-MBBR
To specify Output Channel
To specify Input Port
5-48