To Define Test Specification
In the Test Parameters Definition area:
•Select the Background that will be displayed on the application test setup area of the main
screen. Usually it is the image that shows the device connections.
•Set the Name, data type, Default value, and Description for the parameter. When you add a
parameter, you will see the entry fields at the Properties area. The area shows you the properties
(minimum value, maximum value, number of effective digits, resolution, and unit) for the device
parameter or the test parameter specified by the radio button put on the left of the parameter
name.
•Set the X, Y, and Width values to specify the position of entry field displayed on the application
test setup area.
•Check the Aligh box to set the X origin of entry field to the left edge of the entry field. Uncheck
this box to set it to the left edge of the parameter name placed left next to the entry field.
•Check the Ext box. The entry field will be put on the Extended Test Parameters dialog box that is
displayed by clicking the Extended Setup button on the application test setup area of the main
screen.
Define test parameters
6-7
Set properties
of parameter