To Set Timing Parameters
Measurement Setup
At the Timing area, you set the source output and measurement timing.
•Hold: Source output hold time. Time after the measurement trigger until starting delay time.
•Delay: Measurement start delay time. Time after the hold time until starting measurement.
When a SMU forces pulse, the delay time is ignored. Source output and measurement timing will
be controlled by the pulse width and pulse period.
Trigger
5-11
Measurement time
Delay time
Hold time