Operators Manual
Configure Torque Sensor
The torque can be measured by means of force transducers or a torque-measuring shaft.
The signal is transferred via a ±10 V AC output or a frequency output. In line M, (for
example, motor 1: M1), adjust the following settings:
Column
Gain
Unit
Zero
Unit
Configure Speed Sensor
Possible speed sensors include the incremental encoder (measuring with TTL / AC output)
or an analog signal. In line n (for example, motor 1: n1), adjust the settings:
Column
Gain
Unit
Zero
Unit
Configure Motor or Generator
The Power Analyzer can be used for the analysis of both motors and generators. To
configure the device, adjust the settings in line Drv1 for motor 1:
Column
Type
PPairs
Pref
9-6
Settings
1...
Nm/Hz
Nm/V
1...
Hz, V
1. Move the cursor to a field in line M1 and press Enter.
A list of possible options is displayed.
2. Select a value and press Enter to confirm.
The value is now shown in the display field.
Settings
1...
pul/r
rpm/V
1...
Hz, V
1. Move the cursor to a field in line n1 and press Enter.
A list of possible options is displayed.
2. Select a value and press Enter to confirm.
The value is now shown in the display field.
Settings
MOT
GEN
1 ... 999
P ... P3
Description
Slope
Depending on force transducer or sensing shaft type
Voltage or frequency corresponding to speed = 0
Unit for zero, depending on sensor type
Description
Slope
Pulses per revolution
Revolutions per volt
Voltage or frequency corresponding to speed = 0
Unit for zero, depending on sensor type
Description
Motor
Generator
Number of pole pairs
Reference power for efficiency calculation