Measurement with Voltage and Current Transducer

Figure 5-7 illustrates the connections for measuring the Aron circuit with a voltage and
current transducer.
To prevent damage to the transducer due to overload, check
transducer rating.
Transducer errors limit the measuring bandwidth and reduce the intrinsic
uncertainty.
To avoid possible electric shock or personal injury:
Risk of injury when touching connections, internal circuits
and measuring devices that are not connected to earth
ground.
Always adhere to the instructions regarding the sequence of
connection (see Chapter 5, "Connecting Sequence").
W Caution
Note
XW Warning
Connection to Circuits
Aron Circuit (2-Wattmeter Method, W2)