Page 2
Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.
Table of Contents Title Page Introduction ......................1 How to Contact Fluke ..................1 Safety Information ..................... 2 Hazard Location Information/Approvals ............2 Special Conditions for Safe Use ..............3 Symbols ......................3 Specifications ..................... 4 Available Input Ranges ................. 4 Accuracy 700G Ranges .................
Page 5
List of Tables Table Title Page Symbols ........................3 Equipment Required for Verification ..............9 Verification Points (In psi) ..................10 Calibration Points ....................19 Commands ......................21 Measurement Units Used with Serial Port Commands .......... 22 Error Codes ......................23 Replacement Parts and Accessories ...............
Introduction The 700G Series Pressure Gauges (the Product) are high-accuracy digital pressure test gauges. Accurate to 0.05 % FS, the Products can be used as a calibration reference or in any application where high-accuracy pressure measurement is necessary. The Product features user-configurable functions that include: •...
700G Series Calibration Manual Safety Information A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test. Warning To prevent possible electrical shock, fire, or personal injury: •...
If the Product is exposed to overpressure or sudden physical shock (such as being dropped) examine it for any damage that can cause a safety concern. If necessary, return the Product for evaluation to Fluke. Refer to the How to Contact Fluke section. Warning...
700G Series Calibration Manual Specifications Available Input Ranges See Ranges and Resolution table for available ranges in psi plus equivalent ranges and resolution for all engineering units. Accuracy 700G Ranges Positive Pressure ............±0.05 % FS Positive Pressure (700G01, 700G02) ....... ±0.1 % FS Vacuum ..............
Pressure Gauge Specifications Electromagnetic Compatibility ......... Applies to use in Korea only. Class A equipment (Industrial Broadcasting & Communication Equipment) [1] This Product meets requirements for industrial (Class A) electromagnetic wave equipment and the seller or user should take notices of it.
700G Series Calibration Manual How to Change the Batteries Warning To prevent possible electrical shock, fire, or personal injury, batteries must only be changed in an area known to be non- hazardous. Explosion hazard. To change the batteries, see Figure 1: 1.
Unit Under Test (UUT) is out of tolerance and must be calibrated or repaired. For Product support, see the “How to Contact Fluke” section. The 700G01 and 700G02 require a low-pressure standard such as the Ruska 7250LP.
700G Series Calibration Manual Calibration Adjustment Calibration adjustment is done electronically by internal software with the Product case closed. The recommended adjustment procedure is a 2-point adjust that adjusts a low and high point. This adjustment can be applied to any of the 700 family of gauges. It is the only adjustment that may used with a 700RG ranges.
Pressure Gauge Calibration Adjustment 1. Connect the PC to the Product with the interface cable that comes with 700G/TRACK software. 2. After all connections have been made, vent the Product to atmosphere and send this command: OFFSET_ADJ? 3. Note the value returned. 4.
Page 26
700G Series Calibration Manual push to continue. As the Product measures, the screen will show ———-. When measurements are complete the screen will show the calibration value. 4. Use the Pressure Standard to output 312.50 psi (62.5 %). After the output is stable, push ...
Page 27
Pressure Gauge Calibration Adjustment Table 4. Calibration Points Model 700G01 700G02 700G04 700G05 700G05 700G27 700G07 700G08 Number Pressure Range (psi) 1000 Calibration Point 1 0.3600 1.0000 15.000 30.000 100.00 300.00 500.00 1000.0 Calibration Point 2 0.2880 0.8000 13.000 25.500 87.50 262.50 437.50...
Pressure Gauge Serial Interface List of Commands Table 5 shows the commands that the Product uses to communicate. Send these commands from the PC to set up the Product or to take a measurement. Table 5. Commands Command Description CAL_START Puts the calibrator in calibration mode *CLS Deletes the error queue...
700G Series Calibration Manual Parameter Units The Product can be set to show the measurement units in Table 6. Table 6. Measurement Units Used with Serial Port Commands Units Description Pressure in pounds per square-inch Pressure in bars mBar Pressure in millibars...
Pressure Gauge Error Codes A list of error codes are in Table 7. Table 7. Error Codes Error Description A non-numeric entry was received where a numeric entry is necessary Too many significant digits entered Invalid units or parameter value received Entry is above the upper limit of the allowable range Entry is below the lower limit of the allowable range A required command parameter was missing...
Replacement Parts and Accessories Table 8 lists the customer replaceable parts. Replacement parts can be ordered from Fluke Corporation and its approved representatives. Use the part number when you order the replacement part or accessory. See the “How to Contact Fluke” section.