Introduction; Prior To Measuring; Measuring With Default Configuration; Measuring With User-Defined Configuration - Fluke NORMA 4000 Operator's Manual

Power analyzer
Hide thumbs Also See for NORMA 4000:
Table of Contents

Advertisement

Introduction

The Fluke NORMA 4000/5000 Power Analyzer is designed for the measuring of currents
and voltages for up to three different channels. The Power Analyzer calculates rms values,
real, apparent and reactive power, and other derived values. The accuracy is not affected
by the wave form, frequency, or phase shift. Harmonics may be displayed to a maximum
of half the sampling frequency.
Default settings are available for loading quick setups. In addition, you can define more
specific settings to save and load as required (see, "Configuration" in Chapter 7). The
Power Analyzer begins a measurement as soon as the configuration is set and the device
is switched on.

Prior to Measuring

Connect Power Analyzer to the mains (power) socket.
1. Check the measuring connections at the Power Analyzer.
2. Switch on the Power Analyzer.

Measuring with Default Configuration

If the default configuration is acceptable, no additional steps are required.
Ensure that the factory configuration is loaded (see, "Load Configuration" in Chapter 7).

Measuring with User-Defined Configuration

If custom analysis is required, load the respective configuration (see "Configuration" in
Chapter 7).
If measurements with an external shunt or probe are required, make sure
that there is no signal connected at the direct current inputs. Signals on
both inputs (external- and direct-current input) can damage the
measurement unit.
Note
8
Measuring Process

Introduction

8-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Norma 5000

Table of Contents