Press measuring key WAV three times. The rms values for power in channels 1 through 3
are displayed.
The numbers in subscript for U or I (in the example, U
channel.
As the Power Analyzer requires a complete voltage and current cycle for an
accurate measurement; a full period is automatically added to the average
time of 300 ms of configuration 1:W3, and the new average time is
displayed (for example: 309.9 ms at 22.585 Hz, corresponding to seven
periods).