3-Phase Measurement (W3); Direct Connection - Fluke NORMA 4000 Operator's Manual

Power analyzer
Hide thumbs Also See for NORMA 4000:
Table of Contents

Advertisement

3-Phase Measurement (W3)

Direct Connection

Figure 5-8 illustrates a direct connection for a 3-phase measurement. Ensure that there is
no overload at the current input of the Power Analyzer. If there is a potential risk of
overload at the current input, incorporate a shunt or transducer into the circuit. If
necessary, install appropriate fuses.
To select the 3-phase (W2) method, go to General Setup and select W3. For more
information, see Chapter 7, "Configuration."
To avoid possible electric shock or personal injury:
Risk of injury when touching connections, internal circuits
and measuring devices that are not connected to earth
ground.
Always adhere to the instructions regarding the sequence of
connection (see chapter 5, Connecting Sequence).
L1
L2
L3
N
I
3
Figure 5-8. 3-Phase Measurement-Direct Connection
XW Warning
I
I
2
1
HI
HI
VOLTAGE
VOLTAGE
1000 V max
1000 V max
LO
LO
3
2
PROBE
PROBE
EXT.SHUNT
EXT.SHUNT
10 V max
10 V max
HI
HI
CURRENT
CURRENT
10 A max
10 A max
LO
LO
ALL INPUTS MAX 1000V CATII TO
V
V
V
3
2
1
Max. 10 m
1
Connection to Circuits

3-Phase Measurement (W3)

Load
esn015.eps
5
5-13