A-5 Color Measurement Instruments
6. Turn on the computer.
7. Calibrate the X-Rite DTP41 using ColorWise Pro Tools (see the following section).
8. Calibrate the Fiery X3e using ColorWise Pro Tools and the DTP41 (see

Calibrating the DTP41

For the best color accuracy, you should calibrate the X-Rite DTP41 spectrophotometer
every time you calibrate the Fiery X3e. You can initiate the calibration sequence in
ColorWise Pro Tools Calibrator as part of the process of Fiery X3e calibration. The
computer may also prompt you that it is necessary to calibrate the spectrophotometer.
Alternatively, you can initiate the DTP41 calibration process using the Instrument
button on the DTP41. For more information on this method, see the documentation
that came with the X-Rite DTP41.
Multiple users can be connected to one Fiery X3e server with ColorWise Pro Tools,
but only one user at a time can use the Calibrator module. An error message appears if
you try to calibrate when another user is already using ColorWise Pro Tools to
calibrate.
You will need the Color Reflection Reference strip that is included with the
spectrophotometer. Remove the strip from its protective envelope, handling it by the
edges only. Keep the strip free of dust and smudges and always store it in the protective
envelope.
T
X-R
O CALIBRATE THE
ITE
1. Connect the spectrophotometer to the computer and turn on power (see
2. Start Calibrator.
For instructions on starting Calibrator, see
3. Choose DTP41 as the measurement method.
DTP41
page
2-6.
page
2-10).
page
A-2).