2-17 Using the EFI Densitometer ED-100
Calibrating the Fiery X3e with the ED-100
You can use ColorWise Pro Tools in conjunction with the ED-100 to calibrate
the Fiery X3e.
T
F
O CALIBRATE THE
IERY
1. Start Calibrator.
For instructions on starting Calibrator, see
2. Choose EFI Densitometer ED-100 as the measurement method.
3. In the Check Print Settings pane, choose the desired calibration set.
Choose the appropriate calibration set for the type of media you will use most often.
For this calibration to take effect, the calibration set must be associated with one
N
:
OTE
or more output profiles. There is no need to make any new associations for the default
calibration set, since it is already associated with the default output profile.
4. In the Generate Measurement Page pane, click Print.
The Print Options dialog box appears.
5. Choose the appropriate options and click Print.
For Page Type, choose either 34 or 21 Sorted Patches, or 34 or 21 Randomized
Patches. Sorted patches are printed on the page in the order of their hue and saturation
levels. Randomized patches are printed on the page in random order, to help
compensate for density inconsistencies on different areas of the page.
X3
ED-100
E USING THE
page
2-6.