Calibrating the Libraries in a Multi-Unit Configuration

Figure 52 Multi-Unit
Screen
The
Calibrate All PTMs
mechanisms in all libraries within the multi-unit configuration. Calibrate
the libraries during the initial installation and after any maintenance
procedure.
To calibrate all library PTM's:
1 From the Master library in the multi-unit configuration, press the
tab.
Multi-Unit
The GUI displays the
2 Press the
Calibrate All PTMs
All libraries must be in standby mode.
Note:
The GUI displays a
calibration process is in progress.
Multi-Unit calibration MUST be executed from the master
Note:
library designated during multi-unit configuration.
About
System
Multi-Unit
Off-line
Standby
Load Port
O
Stop
Quantum P4000 and P7000 Libraries User's Guide
Calibrating the Libraries in a Multi-Unit Configuration
command enables you to calibrate all pass through
screen (see
Multi-Unit
button.
Command In Progress
Back
Forward
Operator
Service
Configure
Multi-Unit
Calibrate
All PTMs
Chapter 5 Multi-Unit Commands
figure
52).
dialog box while the
Home
Multi
Unit
5
89