5.2
Environmental conditions of operation.
Ultra-Low Temperature equipment are designed to be safe under the following conditions.
Indoor use
Altitude up to 2000 m (795 mbar)
Temperatures from 10ºC to 30 ºC
Maximum Relative Humidity: 60 %
Voltage fluctuations in mains of up to ±10 % of the nominal voltage.
5.3

Environmental conditions for transportation and storage

Ultra-Low Temperature equipment are designed to be safe under the following transportation
conditions.
Storage temperature: from -15 °C (5°F) to 55 °C (131°F)
Relative humidity: 20 - 85 % (non-condensing)
5.4

Unpacking

The devices are shipped from the factory on a wooden pallet and are packed in sturdy wooden boxes.
The box is screwed to the wooden base. The screws must be removed prior to unpacking to avoid damage to the
unit.
All packaging materials are environmentally friendly and should be reused or recycled. Actively
contribute to the protection of the environment by demanding recyclable packaging and environmentally
friendly disposal methods.
WARNING: Infrico does not recommend knocking the unit forward, sideways or backward. However,
if this occurs, you must ensure that the unit remains in an upright position for at least 24 hours
before connecting it, so that the compressor oil returns to the compressor.