R02
1
R03
60
R04
200
R05
0
R06
0
R07
100
R08
-40
R09
-86
R10
20
PARAM.
DEFECT
MIN.
F01
0
F03
5
F05
0
F07
4
F09
100
F10
0
PARAM.
DEFECT
MIN.
G01
300
PARAM.
DEFECT
MIN.
ES1
24
ES2
24
- Compressing assets if the Tº camera ≥ Setpoint + R01
- Compressing I deactivate if the Tº camera ≤ Setpoint
Minimum time interval between two successive ignitions of
0
30
min
the compressor
0
300
s
Compressor activation delay due to regulator activation
0
300
s
Minimum time between compressor start-up and shutdown
Compressor ON time with compressor use alarm or chamber
0
300
min
alarm
Compressor OF time with compressor use alarm or chamber
0
300
min
alarm
0
100
%
Maximum daily limit of compressor use
-50
50
ºC/ºF
Maximum Allowable SetPoint Value
-50
50
ºC/ºF
Minimum Allowable SetPoint Value
0
100
s
Second compressor activation delay
MAX.
M.U.
Enables evaporator fans
0 = Disabled
0
1
---
1= Enabled
Referring also to the parameter C09
Setpoint ventilators evaporating during on-speed operation
5
50
ºC/ºF
- activated ventilators if evaporating Tº≤ F03
- deactivated ventilators if evaporating T°≥ F03 + 5°C
Evaporator fans during defrost
0
1
---
0 = Disabled
1 = Enabled
Temperature of activation of the ventilators of the evaporator
-40
50
ºC/ºF
after defrost or the ignition
0
300
min
Fan OFF time with compressor shut down
0
300
min
Fan ON time with compressor shut down
MAX.
M.U.
Buzzer reactivation after silencing
0
900
s
0 = Disabled
MAX.
M.U.
Energy saving start time
0
24
h
24h = Excluded
Energy saving end time
0
24
h
24h = Excluded