5.5
Electrical Connection
WARNING!: Connect the equipment to a dedicated outlet with the correct voltage for the device.
Incorrect power or voltage fluctuations can cause serious damage to the equipment.
The equipment is prepared for a feeding of the mains 220-230V 50 Hertz, 220-230V 50 Hertz, 115V,
60Hz. The equipment counts on installed hose and pin in factory. Review the adhesive located in the
feeder. If it does not count on the taking of suitable current, must install it previously. The average one
to isolate the equipment of the feeding network is the pin of the feeder.
WARNING!: This unit must be grounded before use in order to ensure personal safety and the proper
operating of the equipment. A grounding fault may cause damage to personnel or equipment.
Always comply with the National Electrical Code. Do not connect the equipment to power lines that
are already overloaded.
OBLIGATION: The device must be connected to an exclusive dedicated circuit. Failure to comply with
this requirement shall void the warranty.
WARNING!: The device is designed to tackle a voltage fluctuation of around 10 % in relation to the
rated voltage indicated in the rating plate. A compressor fault due to higher fluctuations shall
automatically void the warranty.
Note: It is recommended to install a UPS (Uninterruptible Power Supply) or other system to avoid
voltage peaks or lack of electric current supply.
WARNING!: If the hose or the peg are altered in any way, they may constitute a serious hazard. Any
alteration of these components shall void the warranty.
WARNING: Devices connected to an extension cord are not covered by Infrico's warranty.
WARNING: The power cable can only be replaced by an authorized technical support service
provider.