Locking/Unloking The Keyboard; Switching Interior Light On/Off; Mute Alarm - Infrico Lab Care Plus ULF70086 Service Manual

Ultra low temperature freezer

Advertisement

6.6

Locking/Unloking the Keyboard

The keyboard lock is enabled. The lock can be enabled manually from the HOME screen by pressing the
HOME and MENU keys successively. Next, a closed padlock icon will be displayed for 3 seconds.
The keyboard lock prevents access to all menus and functions except for the following ones:
- Switching light on/off
- Mute audible alarm
To unlock the keyboard, press the
icon
will be displayed for 3 seconds.
The procedure is facilitated by a message that is displayed each time a key is pressed, and the keyboard
is locked.
6.7

Switching interior light on/off

This function can be accessed manually, even if the keyboard is locked. The light switches on after
pressing the 'light' key and after opening/closing the door, enabled in parameter C01.
6.8

Mute Alarm

To disconnect the audible alarm, proceed as follows:
1.- Make sure that no process is in progress.
2.- Press any key (when any key is pressed, the corresponding function is not enabled)
When pressing the key, the following icon
audible alarm is re-enabled. If G01=0, the audible alarm will not be enabled.
HOME and MENU keys
will be displayed and after G01 seconds the
14
Service Manual and Warranty
successively: An open padlock