8.7
List of Parameters
PARAM.
V.DEFECT
MIN.
A01
-10
A02
10
A03
300
A04
60
A05
-20
A06
45
A07
0
A08
0
A09
0
PARAM.
DEFECTO
MIN.
D01
0
D02
1
D03
1
D06
-
D07
1
D08
5
MAX.
M.U.
Differential of alarm of low temperature in camera (relative to
-50
0
ºC/ºF
0
50
ºC/ºF
Differential of alarm of high temperature (relative to setpoint)
Retardation of alarm of discharge and low temperature from
0
300
min
0
900
min
Minimum allowable evaporator temperature (relative to set
-50
0
ºC/ºF
0
90
ºC/ºF
0
60
ºC/ºF
0
20
ºC/ºF
0
20
ºC/ºF
MAX.
M.U.
Temperature measurement unit
0
1
---
0 = Celsius
1 = Fahrenheit
Display of the initial screen of welcome
0
1
---
0 = none
1 = INFRICO
Backlight
0 = ignited pressing any key, after minto, the bottom light is
0
1
---
extinguished and only key HOME is illuminated.
1 = Always ignited.
Blockade of screen of the temperature of camera during
defrost.
0
1
---
0 = no blocking
1 = enabled
1
300
min
Time out for keyboard lock
Keyboard lock
0 = no authorised
0
2
---
1 = blockade/unfreezing allowed by key
2 = lock activated per time of delay or key, opening allowed
by key
Service Manual and Warranty
ALARMS
setpoint)
the ignition of the device or finalized defrost
Alarm delay high and low temperature.
Maximum allowable condenser temperature (relative to set
Temperature differencial for dirty condenser alarm
Low temperature alarm hysteresis (relative to A01)
High temperature alarm hysteresis (relative to A02)
ALARMS
31
point)
point)