6.9
`User Identification´ Function
The 'user identification' function can be enabled/disabled using the ID parameter
It is possible to programme up to 6 user codes and a single user 1 (default code 111) with permission to
set and modify access credentials for all other users. If the password ID 1 is forgotten, use code 666 to regain
access.
The controller requires the user identification code to perform any of the following actions:
- Turn the device on/off
- Edit the set point
- Manual defrost
- Access to parameters
- Unlock the door after electronic locking
The actions performed will be recorded as special events in the historical data, along with the indication
of the user who performed them. Each event will make a recording of the relevant temperatures and set point,
which guarantees the continuity of the historical record results.
Find below an example of how to enter the user code when turning on the device:
- With the device in stand-by mode, press the ON/STAND BY key
- When the user login screen appears, enter the correct code and press the SET key to confirm and
access the device.
Similarly, the same user identification procedure will be required to turn off the device.
After logging in, any action performed within one minute of entering the login code will not require user
information validation.
6.10

Battery Operation during a Power Outage

The controller has a backup battery connected to an EVC99P00 battery charger. While there is battery
charge, it is possible to save data related to the product probes and alarms.