ES3
2
PARAM.
DEFECT
MIN.
PR1
30
PR2
1
PARAM.
DEFECT
MIN.
ADD
247
MB1
2
MB2
2
PARAM.
DEFECT
MIN.
RLA
1
PARAM.
DEFECT
MIN.
AUX1
-10
AUX2
10
AUX3
2
AUX4
2
AUX8
60
AUX9
300
AUX10
-
I0
0
ID
1
CNF01
3
0
20
ºC/ºF
SetPoint increase during energy saving time
MAX.
M.U.
1
60
min
Recording Time
USB
0
1
---
0 = Disabled
1 = Enabled
MAX.
M.U.
1
247
---
Instrument address
Baut rate
0 = 2400
0
2
---
1 = 4800
2 = 9600
Parity
0 = no parity
0
2
---
1 = Odd
2 = Even
MAX.
M.U.
0 = Disabled if alarm in progress
0
1
---
1 = Enabled if alarm in progress
MAX.
M.U.
Differential of alarm of low temperature in product (relative
-50
0
ºC/ºF
to setpoint)
Differential of alarm of high temperature in product (relative
0
50
ºC/ºF
to setpoint)
0
20
ºC/ºF
Low temperature alert hysteresis (relative to AUX1)
0
20
ºC/ºF
High temperature alert hysteresis (relative to AUX2)
0
300
min
High or Low temperature auxalarm delay
0
300
min
Aux temperature alarm delay due to regulator start-up
Main Vista in screen
0
1
---
0 = Camera probe
1 = Product probe
0
1
---
Display contrast
Usuary code identification
0
1
---
0 = deactivated
1 = activated
Name selection for regulation probe on power board
0
3
---
0 = Camera probe
Service Manual and Warranty
ALARMS
ALARMS
ALARMS
ALARMS
36
Need help?
Do you have a question about the Lab Care Plus ULF70086 and is the answer not in the manual?