2.2 Temperature Monitoring
Important note: We recommend the use of a redundant and independent temperature monitoring
system, so that the ultra low freezer can be continuously monitored for a performance commensurate
with the value of the stored product.
2.3 Initial Charge
Allow the freezer to operate at the desired temperature for a minimum of 12 hours before loading.
Load the ultra low freezer from a shelf on a shelf, starting with the top shelf. After loading on each shelf,
allow the ultra low freezer to recover to desired set point before loading the next shelf. Repeat this process until
the freezer is fully charged.
Caution Failure to follow these procedures or overloading the unit can cause excessive stress on the
compressors or jeopardize the safety of the product.