NOTE:
•
• If the scanner's firmware has been updated previously, you must re-calibrate
16
Spector-AF10 User's Manual
The calibration process may take quite a time. Please be patient for the
finishing of the process.
your scanner to ensure optimal scanning results. Before re-calibrating, you
should remove the existing calibration data from your system using one of
the following methods (A or B):
A. Open the following path <C:\Program Files (x86)\Microtek\ScanWizard
DI\Scanners>, where ScanWizard DI software program is saved. Find
and remove "Calidata601f_xxxxxxxxxxx.dat" and "Calidata601f_
xxxxxxxxxxx.in"
files from your system.
Once the calibration data is removed, please restart the ScanWizard DI
software. You will be prompted to re-calibrate your scanner. Follow the
instructions in the prompted window and refer to the previous page to
calibrate your scanner.
B. Reset all calibration data:
1) Launch the ScanWizard DI software and click the Scan Setup button
on the main interface to access the Scan Setup interface.
2)
Click the Information button (
interface, then select "Scanner Maintenance" from the options
menu.
3) In the opened window, click the Execute button next to the "Re-
calibrate UV" option.
You will be prompted to re-calibrate your scanner once more. Follow
the instructions in the prompted window and refer to the previous
page for calibration guidance.
When calibration is completed, the previous calibration data located
at the path <C:\Program Files (x86)\Microtek\ScanWizard DI\
Scanners> will be replaced with the new calibration data.