Quick Reference
Quick Reference
Table 14-1.
Routine maintenance tasks
14-2
The following chart lists routine maintenance tasks, when you perform
each task, and the equipment on which you perform them.
When
Equipment
Daily or when used
CUSA EXcel
console
Daily or when used
Cooling water
reservoir
Daily or when used
Footswitch
Every 3 months, or every 12
CUSA EXcel
to 15 procedures, whichever
console
comes first
Every 6 months, or when
CUSA EXcel
needed
console
Once a year
CUSA EXcel
console
Every month or 80 hours
Handpiece
of use
Every 6 months or 500 hours
Handpiece
of use
CUSA Excel Ultrasonic Surgical Aspirator System User's Guide
Task
Disinfect and wipe dry.
Discard water from the reservoir.
1. Discard the footswitch cover.
2. Clean the footswitch.
Fill the cooling water reservoir with a solution
consisting of 100 cc of 70% alcohol and 900
cc of water. Connect the handpiece, then turn
on the system. Allow the system to run a few
cooling cycles to clean the interior tubing.
Turn off the system. Discard the alcohol
solution, then fill the reservoir with distilled
water. Turn on the system, and allow it to
cycle the water to remove the alcohol solution
from the tubing.
Replace the contamination guard. Write either
the installation date or the expiration date on
the guard.
Contact a Integra service representative to
replace the cooling water pump tubing.
Apply a high viscosity lubricant to the
handpiece connector o-rings.
Replace the handpiece connector o-rings.