Prepare The System For Startup; At The Rear Panel; On The Side Panel: Cooling Water - Integra cusa excel User Manual

Ultrasonic surgical aspirator system
Hide thumbs Also See for cusa excel:
Table of Contents

Advertisement

Prepare the System for Startup

CUSA Excel Ultrasonic Surgical Aspirator System User's Guide

At the Rear Panel

Plug the CUSA EXcel system power cord into a wall receptacle.
1.
Confirm that the AC Main Switch on the rear panel is on ( | ).
2.
Verify that the CUSA EXcel system footswitch is connected to the rear
3.
of the console.
Follow hospital policies and procedures regarding placing
4.
footswitches in plastic bags.

On the Side Panel: Cooling Water

Remove the cooling water reservoir from the console:
1.
Press the button (just above the reservoir) on the Arm Housing to
a.
release the fittings inside the housing.
Slide the reservoir toward the rear of the console, then remove it
b.
from its slot.
Gently open the black rubber lid on the reservoir top.
2.
Fill the cooling water reservoir to the line with 1000 cc distilled water
3.
(sterile water is distilled).
Gently close the black rubber lid.
4.
Slide the cooling water
5.
reservoir into the slot on the
console side panel until it
snaps into place.
Prepare the System for Startup
off.
Hz 900
110-120/220-240V~
50/60
SN
VA
with
intermittent
operation
loading,10
Min.
on, 5
Min.
Inc,
Classification:
Mfg.
BF
Type
Center,
Date
CO 80301-3299
Boulder,
Services
S.A.
Continuous
USA
Valleylab
Auto
Suture
European
78990
Elancourt,
Made
in USA
France
5-7

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the cusa excel and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents