Prepare the System for Startup
CUSA Excel Ultrasonic Surgical Aspirator System User's Guide
At the Rear Panel
Plug the CUSA EXcel system power cord into a wall receptacle.
1.
Confirm that the AC Main Switch on the rear panel is on ( | ).
2.
Verify that the CUSA EXcel system footswitch is connected to the rear
3.
of the console.
Follow hospital policies and procedures regarding placing
4.
footswitches in plastic bags.
On the Side Panel: Cooling Water
Remove the cooling water reservoir from the console:
1.
Press the button (just above the reservoir) on the Arm Housing to
a.
release the fittings inside the housing.
Slide the reservoir toward the rear of the console, then remove it
b.
from its slot.
Gently open the black rubber lid on the reservoir top.
2.
Fill the cooling water reservoir to the line with 1000 cc distilled water
3.
(sterile water is distilled).
Gently close the black rubber lid.
4.
Slide the cooling water
5.
reservoir into the slot on the
console side panel until it
snaps into place.