Communication Interface Specifications

You can use the USB port or RS-232C/422 connector of the Controller to perform serial communication with
external devices such as a personal computer or programmable controller.
Serial communication functions in the RUN mode. Communication cannot be performed in the ADJ or MENU
modes.
<USB>
This interface allows Full speed (12 Mbps) communications compliant with USB2.0 with a PC equipped with
the same USB interface.
Synchronization method Start-stop
Transmission code
ASCII (Binary format can be selected only when outputting measurement values set at
[Setup] - [Support] - [Calculation] - [Data].)
Communication speed
USB2.0-compliant
Data length
-
Parity
-
Stop bit
-
Delimiter
CR, LF, CR+LF
Transmission protocol
Normal (Note, however, that XMODEM protocol is used when sending image data, system
data and other data.)
<RS-232C/422>
This interface allows data communications compliant with the EIA RS-232C/422 standards up to a maximum
speed of 115200 bps.
Synchronization method Start-stop
Transmission code
ASCII (Binary format can be selected only when outputting measurement values set at
[Setup] - [Support] - [Calculation] - [Data].)
Communication speed
9600, 19200, 38400, 57600, 115200
Data length
7 bits, 8 bits
Parity
None, even, odd
Stop bit
1 bit, 2 bits
Delimiter
CR, LF, CR+LF
Transmission protocol
Normal (Note, however, that XMODEM protocol is used when sending image data, system
data and other data.)
For details on how to set the communication specifications, refer to the User's Manual.
<Ethernet>
Communication protocol TCP/IP
Transmission mode
Peer to Peer