Introduction; Safety Information; Operation Safety Precautions - Labnet Prism C2500-100V Instruction Manual

Microcentrifuge
Table of Contents

Advertisement

1.0 Introduction

The Labnet Prism™ microcentrifuge is a small benchtop centrifuge designed for separation of various research samples. The motor is
brushless and requires no routine maintenance. This microcentrifuge is supplied with a 24 x 1.5/2.0 mL rotor for microsample tubes.
Adapters are available for tubes smaller than 1.5 mL. This reaches speeds of up to 15,000 rpm/21,200 x g.

2.0 Safety Information

Before using the Labnet Prism microcentrifuge for the first time, please read this entire manual carefully. If the equipment is used in a
manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
This symbol indicates a potential risk and alerts you to proceed with caution.

2.1 Operation Safety Precautions

Never use the centrifuge in any manner not specified in these instructions.

Never operate the centrifuge without a rotor properly attached to the shaft.

Always use a wrench to tighten the rotor nut. Never tighten rotor nut by hand only.

Never fill tubes while they are in the rotor. Liquid spillage may harm unit.

Never put hands in the rotor area unless the rotor is completely stopped.

Never move the centrifuge while the rotor is spinning.

Never use solvents or flammables near this or other electrical equipment.

Never centrifuge flammable, explosive, or corrosive materials.

Never centrifuge hazardous materials outside of a hood or a proper containment facility.

Always load the rotor symmetrically. Each tube should be counter-balanced by another tube of the same type and weight.

Always locate the centrifuge within easy access to an electrical outlet.

Always use only microcentrifuge tubes made from plastic and designed to withstand centrifugal forces of at least 21,200 x g.

Always use a wrench to tighten the rotor nut.

Do not operate the centrifuge if any of the following conditions exist:
The centrifuge has not been installed properly.

The centrifuge is partially dismantled.

Service has been attempted by unauthorized or unqualified personnel.

The rotor has not been installed securely on the motor shaft.

Rotors and accessories outside of the standard range are not to be used without permission from the manufacturer.

Exception: Microcentrifuge tubes made of plastic, normally available in the laboratory.
The centrifuge is located in an explosive atmosphere.

Materials to be centrifuged are combustible and/or explosive.

Materials to be centrifuged are chemically reactive.

The rotor load is not properly balanced.

The rotor nut was not tightened with a wrench.

Instruction Manual
|
1