1.0 Introduction
The Labnet Spectrafuge™ 24D Microcentrifuge is a small benchtop centrifuge designed for separation of various research samples.
The motor is brushless and requires no routine maintenance. The Spectrafuge 24D is supplied with an 24 x 1.5/2.0 mL rotor for micro-
samples. Adapters are available for tubes smaller than 1.5 mL. The Spectrafuge 24D reaches speeds of up to 13,300 rpm/16,300 x g.
2.0 Symbols and Conventions
The electrical warning indicates the presence of a potential hazard which could result in electrical shock.
CAUTION: This symbol refers you to important operating and maintenance (servicing) instructions within this Instruction
Manual. Failure to heed this information may present a risk of damage or injury to persons or equipment.
This symbol identifies a Protective Earth (PE) terminal, which is provided for connection of the supply system's protective
earth (green or green/yellow) conductor.
3.0 Safety Information
NOTE: All users of the centrifuge must read the Safety Information section of this manual before attempting to operate the unit.
If this equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be
impaired.
Never use the centrifuge in any manner not specified in these instructions.
Never operate the centrifuge without a rotor properly attached to the shaft.
Never tighten rotor nut by hand only, use a wrench.
Never fill tubes while they are in the rotor. Liquid spillage may harm unit.
Never put hands in the rotor area unless the rotor is completely stopped.
Never move the centrifuge while the rotor is spinning.
Never use solvents or flammables near this or other electrical equipment.
Never centrifuge flammable, explosive, or corrosive materials.
Never centrifuge hazardous materials outside of a hood or proper containment facility.
Always load the rotor symmetrically. Each tube should be counterbalanced by another tube of the same type and weight.
Always locate the centrifuge within easy access to an electrical outlet.
Always use only microcentrifuge tubes made from plastic and designed to withstand centrifugal forces of at least 16,500 x g.
Do not operate the centrifuge if any of the following conditions exist:
The centrifuge has not been installed properly.
The centrifuge is partially dismantled.
Service has been attempted by an unauthorized or unqualified personnel.
The rotor has not been installed securely on the motor shaft.
Rotors and accessories not belonging to the standard range are being used without permission being obtained from the
manufacturer to use such rotors and/or accessories in the centrifuge. Exception: Microcentrifuge tubes made of plastic, normally
available in the laboratory.
The centrifuge is located in an explosive atmosphere.
Materials to be centrifuged are combustible and/or explosive.
Materials to be centrifuged are chemically reactive.
The rotor load is not properly balanced.
The rotor nut was not tightened with a wrench.
Instruction Manual
|
1