10.0 Determination of G-values
The centrifuging radius of the 1.5/2.0 mL rotor is 8.23 cm. The Labnet Prism™ microcentrifuge has an automatic g-force conversion
program, so g-values are automatically calculated and can be displayed on the control panel, for this centrifuging radius. If adapters
or smaller tubes are used, the centrifuging radius changes as does the g-force. The chart below can be used to manually determine
g-values for any centrifuging radius.