Order List; Annex; Glossary - Dräger Alcotest 7000 med Instructions For Use Manual

Hide thumbs Also See for Alcotest 7000 med:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 15
en
|

Order list

Breath flow
Breath duration
Measurement accu-
racy
Standard deviation of repeatability using standard alcohol
Up to 0.50 mg/L
≥0.50 mg/L
Up to 1.00 ‰
≥ 1.00 ‰
Sensitivity drift
Wait time until next measurement can be taken following
a previous measurement of:
0.00 mg/L 2 s
0.50 mg/L 40 s
1.00 mg/L 80 s
>1.50 mg/L 120 s
Adjustment
Dimensions
Weight
Power supply
Lithium-ion battery
Interface
Classification in accordance with Regulation (EU)
2017/745
Measuring device
Mouthpieces
LED risk group
Protection class
UMDNS code
DIMDI registration
number
CE marking
The device meets the requirements of EN 15964:2011. The
nameplate and the instructions for use differ from the stan-
dard. It can be configured in accordance with the standard.
Display of measured values if configured in accordance with
EN 15964:
Measured value
0.0 mg/L to 0.03 mg/L
>0.03 mg/L
24
>4 L/min (default)
>0.5 s (default)
0.008 mg/L
1.7% of the measured value
0.017 ‰ or
1.7% of the measured value
Depending on which value is
higher
Typically 0.4 % of the mea-
sured value per month
0.00 ‰ 2 s
1.00 ‰ 40 s
2.00 ‰ 80 s
>3.00 ‰ 120 s
Recommendation: every
12 months
58 mm × 170 mm × 32 mm (W × H
× D)
approx. 245 g
Permanently installed in the
device
USB-C
Class I with measuring function
Class I
2
IP 54
17-475
DE/CA01/102/XXX
Electromagnetic compatibility as
per IEC 60601-1 and IEC 60601-
1-2
Display
0.0 mg/L
Measured value
10
Order list
Designation and description
Alcomed mouthpiece with valve (100 pieces)
Wrist strap
Charging cradle
USB-C cable
Power supply unit
Dräger Mobile Printer BT
Standard ethanol solution 1.21 g/L (1 x 500 mL) 6728838
11

Annex

11.1 Glossary

Unit
Explanation
°C
Degrees Celsius; unit for temperature
°F
Degrees Fahrenheit; unit for temperature
hPa
Hectopascal; unit for ambient pressure
% r. h.
Percent relative humidity; relative atmospheric
humidity
h
Unit for hours
mg/L
Milligram per liter; unit for breath alcohol con-
centration
%
Percent; unit for blood alcohol concentration
Per mill; unit for blood alcohol concentration
g/L
Gram per liter; unit for blood alcohol concentra-
tion
Blood
Conversion factor between breath alcohol and
Breath
blood alcohol concentration
Ratio
MDR
Medical Device Regulation
Instructions for use
Order no.
3705618
3700416
3700420
3700418
3701090
3700421
®
|
Dräger Alcotest
7000 med

Advertisement

Table of Contents
loading