Troubleshooting

4.10.15 Changing the PIN
Changes the PIN.
● In the menu, select Login.
● Enter the relevant PIN.
● Press
to confirm.
● Select the Change PIN menu item using the
buttons.
● Press
to confirm.
● Enter the new PIN.
● Press
to confirm.
4.10.16 Turn off
Turns off the device.
● Briefly press the
button.
 The device turns off.

4.11 Printing Results

To print the measurement results, the device can be
connected to the Dräger Mobile Printer BT (see "Order list",
®
page 24) via Bluetooth
.
Observe the Dräger Mobile Printer BT instructions for use.
Alternatively, the measurement results can also be printed
using the Dräger Mobile Printer.
Observe the Dräger Mobile Printer instructions for use.
● In the menu, select Login.
● Enter PIN-1.
● Press
to confirm.
● In the PIN-protected menu, select Printer.
● Under Printing mode, select the Mobile Printer or
Mobile Printer BT.

4.12 Exporting data

The data stored by the device can be exported and then
®
analyzed in Microsoft
Excel, for example. (This function is
available from firmware version 2.0.0).
1. Connect the USB-C cable to the USB-C port of the device
and the USB port of a PC or laptop.
 The device turns on.
2. Log in to the PIN-1 menu in the device menu and choose
Transfer mode.
 "Transfer mode" is shown on the device display.
3. Open Windows Explorer.
 The device is now displayed as a USB drive.
4. Open the path "USB drive\var\scenario_logs\BreathTest".
5. Select the CSV file.
6. Select Copy.
7. Select the desired drive and directory as the storage
location.
8. Select Save.
 The CSV file is saved.
9. Turn off the device by pressing the
Transfer mode.
22
5
An error message is displayed in the event of a device
malfunction. The number below the message is for service
purposes. If the error persists even after turning the device off
and on multiple times, contact authorized personnel, such as
DrägerService.
and
Serious incidents related to the product must be reported to
the manufacturer and the responsible authorities.
Error/Message
Device does not
turn on.
Insufficient volume
of breath/
sample incomplete
Device out of tem-
perature range.
↑ ↑ ↑
(over measuring
range)
Insufficient volume
of breath. [Only for
breath samples
using the mouth-
piece]
Volume of breath
too large. [Only for
breath samples
using the mouth-
piece]
Sample disruption
Error number
Adjustment inter-
val expired
button to exit
Troubleshooting
Cause
The rechargeable
batteries are
empty.
Subject not blow-
ing strongly
enough or blowing
interrupted.
Device is too warm
or too cold.
Measuring range
exceeded. Resid-
ual mouth alcohol
can distort the
measured values.
Volume of breath
sample insufficient.
Measuring range
exceeded.
Irregular exhala-
tion, abrupt cessa-
tion or inhalation.
Device error.
The adjustment
interval has
expired.
Instructions for use
|
Dräger Alcotest
Remedy
Recharge the
rechargeable bat-
teries.
Blow into the
mouthpiece more
strongly and with-
out interruption.
If necessary, carry
out a passive
breath sample.
Allow the device to
cool down or warm
up.
A wait time of at
least 15 minutes
after alcohol was
last consumed
must have
elapsed. Press the
button to repeat.
Repeat the breath
sample after press-
ing the
button.
If necessary, carry
out a passive
breath sample.
Repeat the breath
sample after press-
ing the
button.
Repeat the breath
sample after press-
ing the
button.
If necessary, carry
out a passive
breath sample.
Turn the device off
and then back on.
If the error persists,
contact DrägerSer-
vice or your dealer.
Have adjustment
carried out by
authorized person-
nel or DrägerSer-
vice.
This does not
impair the device
being ready to
measure.
®
7000 med