Menu And Context Menu; Prerequisites For Use; Preparations For Use; Breath Sample - Dräger Alcotest 7000 med Instructions For Use Manual

Hide thumbs Also See for Alcotest 7000 med:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 15
en
|
Use
4.2

Menu and context menu

To access the menu:
● Press the
button.
To access the context menu:
● Press the
button
To navigate in the menu or context menu or to configure the
settings:
1. Press the
or
button.
2. Press the
button to confirm the selected feature/entry.
To exit the menu or context menu:
● Select the Back menu entry.
4.3

Prerequisites for use

CAUTION
Risk of swallowing
Do not distort the mouthpiece in the mouth.
► The user must observe the subject during use.
The device is adjusted in the factory before delivery. Dräger
recommends that the device be adjusted every 12 months by
DrägerService.
● The ambient air must be free of alcohol and solvent
vapors (disinfectants).
● At least 15 minutes must have elapsed since the subject
last consumed alcohol.
● The wait time cannot be reduced by rinsing your mouth
with water or with non-alcoholic beverages.
● Residual mouth alcohol or flavored drinks (e.g. fruit juice),
breath sprays containing alcohol, medicinal juices and
drops can distort the result. Observe a wait time of at least
15 minutes in these cases as well.
● Belching and vomiting can distort the result.
● The device detects exhaled samples with a volume
greater than 0.3 L. The subject must be able to exhale at
least this volume. If the volume of the exhaled sample is
greater than 0.3 L and less than 1.3 L, the message "Low
volume" appears on the display in addition to the
measurement result.
The displayed breath alcohol result only corresponds to
the blood alcohol concentration for samples with a volume
greater than 1.3 L. All breath alcohol tests with the "Low
volume" message present may display a lower result than
the actual blood alcohol concentration.
● For emergency use or other critical measurements,
always have a second functioning device or an alternative
means of measurement available.
● The seal sticker over the name plate on the reverse of the
device must not be damaged.
4.4

Preparations for use

1. Ensure that the rechargeable battery is charged.
2. Turn the device on by pressing the
 An air blank is carried out.
An air blank is a test to see whether alcohol can be
detected in the ambient air, which could distort test
results.
3. If no alcohol is detected, wait until the device displays that
it is ready to measure (see fold-out page, Figure B: 1).
18
4.4.1 Before sampling
● The subject must be breathing calmly and steadily.
● The subject should be able to provide the minimum
4.5
Overview of the different measurement modes:
Automatic
breath sam-
ple
Low-vol-
ume breath
sample
Passive
breath sam-
ple
4.5.1 Taking an automatic breath sample
Use a new mouthpiece for each measurement.
Do not allow any high alcohol concentrations on the sensor,
e.g., by purging the mouth with a high proof alcohol shortly
before the measurement. Doing so shortens the life span of
the sensor.
When removing the mouthpiece (see "Order list", page 24)
from the packaging, ensure that the part that goes into the
mouth remains in the packaging. The mouth contact area
must remain protected for hygiene reasons.
The mouthpiece can be inserted into the mouthpiece
receptacle from both sides (right or left).
1. Open the mouthpiece packaging using the prefabricated
2. Insert the mouthpiece upwards and tilted at the corner of
3. Hold the mouthpiece in the tilted position and slide
4. Press the mouthpiece into the mouthpiece receptacle of
button.
5. Remove the rest of the mouthpiece packaging.
volume of breath (>1.3 L) required for sampling.

Breath sample

Device sta-
Trigger
tus
with mouth-
blow until end
piece
of the breath
with mouth-
permanently
piece
active,
blow until end
of the breath
without
Press the
mouthpiece
button
perforation and remove the part of the packaging that will
be placed on the device. Leave the mouth contact area
protected in the packaging for hygiene reasons (see fold-
out page, Figure A).
the mouthpiece receptacle casing so that the contact
surface (serrated marking) is pointing upwards (see fold-
out page, Figure A, a).
upwards until the mounting spacer stops the movement at
the bottom of the mouthpiece (see fold-out page Figure A).
the device on the contact surface (see fold-out page,
Figure A, b).
 The mouthpiece audibly clicks into place.
If the device displays READY, a measurement or
follow-up measurement can be carried out.
Instructions for use
Result
Measured
value
Measured
value
Notice: "Low
volume"
Alcohol or no
alcohol, mea-
sured value,
notice: pas-
sive (if sec-
ond unit is
disabled)
®
|
Dräger Alcotest
7000 med