Comparing the Control Solution Test Results

The test result of each control solution should be within the
range printed on the label of the test strip vial. Repeat the
control solution test if the test result falls outside of the range.
Out of range results may occur due to the following factors:
Situations
• When the control solution bottle was
not shaken well,
• When the meter, test strip, or the
control solution were exposed to high
or low temperatures,
• When the first drop of the control
solution was not discarded or the tip
of the bottle was not wiped clean,
• When the meter is not functioning
properly.
• When the control solution is past the
expiration date printed on the bottle,
• When the control solution is past its
discard date (the date the bottle was
opened plus three months),
• When the control solution is
contaminated.
If results continue to fall outside the range printed on the
test strip vial, the CareSens N Test Strip and Meter may not be
working properly. Do not use your system and contact your
authorised i-SENS sales representative.
Actions
Repeat the control
solution test by
referring to the Note
on page 21.
Discard the used
control solution and
repeat the test using
a new bottle of
control solution.
25