i-SENS CareSens N Voice User Manual
Hide thumbs Also See for CareSens N Voice:

Advertisement

User's Manual
Blood Glucose Monitoring System
For single user / home use only
No
Coding
Talking
Function
µL
PGA1E3217 REV0 01/2013

Advertisement

Table of Contents
loading

Summary of Contents for i-SENS CareSens N Voice

  • Page 1 User's Manual Blood Glucose Monitoring System Coding Talking Function µL For single user / home use only PGA1E3217 REV0 01/2013...
  • Page 2 Seoul 139-845, Korea www.i-sens.com Contact Customer Service: 1-800-429-5001 9 am - 9 pm EST, Monday - Saturday At all other times or in case of emergency, please contact your healthcare professional or emergency medical response. © 2013 i-SENS, Inc. All Rights Reserved.
  • Page 3: Table Of Contents

    Information Important Information: Read This First! Important Health-Related Information Specifications CareSens N Voice Blood Glucose Monitoring System Inserting or Replacing the Batteries CareSens N Single Blood Glucose Test Strip CareSens N Voice Blood Glucose Meter CareSens N Voice Blood Glucose Meter Display...
  • Page 4: Important Information: Read This First

    • The CareSens N Voice Blood Glucose Meter is designed to after eating, insulin injection or exercise. minimize code related errors in monitoring by using the no-coding function.
  • Page 5: Specifications

    Quick Reference Guide  Test time 5 seconds Batteries (2)  Sample type Fresh capillary whole blood CareSens N Voice System may include the following items: CareSens N Single Blood Glucose Test Strips (10) Calibration Plasma-equivalent  Lancing Device  Assay method...
  • Page 6: Inserting Or Replacing The Batteries

    Inserting or Replacing the Batteries CareSens N Single Blood Glucose Test Strip The CareSens N Voice Meter uses two AAA alkaline batteries. The CareSens N Voice Blood Glucose Monitoring System Before using the meter, check the battery compartment and measures blood glucose quickly and accurately. It automatically insert batteries if empty.
  • Page 7: Caresens N Voice Blood Glucose Meter

    CareSens N Voice Blood Glucose Meter Data Port • Do not reuse test strips. Used to transfer data from the • Do not use CareSens N Single Test Strips beyond the meter to a computer with a expiration date. This may cause inaccurate results.
  • Page 8: Setting Up Your System

    (L2, Spanish), press ▲ or ▼ button. Press the S button to Note: The unit of measurement for your CareSens N Voice confirm the language and set the year. Meter is preset and fixed in mg/dL. 12 www.i-sens.com...
  • Page 9 Press ▲ or ▼ button until the correct blinking number shows the correct date. Press minute appears. After setting the minute, press the S button to confirm the date and set the the S button to set the sound. time. 14 www.i-sens.com www.i-sens.com 15...
  • Page 10: Setting The Hypoglycemia Indicator

    After completing meter setting, press and hold the S button for 3 seconds to confirm all your selections and exit from SET mode. If the talking function is on, the meter will announce the set date and time for your confirmation. 16 www.i-sens.com www.i-sens.com 17...
  • Page 11: Checking The System

    Please apply control solution onto the test strip." You should do a control solution test: • When you want to practice the test procedure using the control solution instead of blood; • When using the meter for the first time; 18 www.i-sens.com www.i-sens.com 19...
  • Page 12 ▲ button for 3 seconds until the ctl appears on the display. When the ctl is displayed, the result is stored in the meter's memory but is not included in the averages. 20 www.i-sens.com www.i-sens.com 21...
  • Page 13: Comparing The Control Solution Test Results

    If results continue to fall outside the range printed on the test • Keep your meter and lancing device clean. See the strip vial, the CareSens N Voice Meter and CareSens N Single pre-cleaning and disinfection section on page 41.
  • Page 14: Preparing The Lancing Device

    The smaller numbers are for a shallower puncture and the larger numbers are for a deeper puncture. Choose a depth of penetration by rotating the top portion of the adjustable tip until the setting number matches the arrow. 24 www.i-sens.com www.i-sens.com 25...
  • Page 15: Flagging Post-Meal Test Results

    Applying the Blood Sample Flagging Post-meal Test Results Step 8 CareSens N Voice Meter allows you to flag a result of a post- Obtain a blood sample using the lancing meal test with symbol. The post-meal flag ( ) can be device.
  • Page 16: Discarding Used Lancets

    • The lancet is very sharp. Please keep away from children. • Keep the lancets in a cool and dry place. Caution: The lancet is for single use only. Never share or reuse a lancet. Always dispose of lancets properly. 28 www.i-sens.com www.i-sens.com 29...
  • Page 17: Alternative Site Testing

    Keep holding the lancing device against your skin • When insulin has the biggest effect until sufficient (at least 0.5 μL) blood is drawn. Carefully lift the • Two hours after an insulin injection lancing device away from your skin. 30 www.i-sens.com www.i-sens.com 31...
  • Page 18: Hi And Lo Messages

    HI and Lo Messages AST Precautions HI Message • Do not ignore the symptoms of hyperglycemia or The CareSens N Voice Meter displays results hypoglycemia. between 20 - 600 mg/dL. HI appears when the blood glucose level is greater than 600 mg/dL •...
  • Page 19: Target Blood Glucose Ranges

    Target Blood Glucose Ranges Meter Memory The CareSens N Voice Meter can save up to 500 test results with Reminders Your target ranges time, date and temperature. If the memory is full, the oldest test from your healthcare professional Time of day result will be deleted and the latest test result will be stored.
  • Page 20: Viewing Test Results

    After checking the stored test result, press the S Step 5 button to turn off the meter. Use the ▼ button to scroll back through the averages seen previously. To turn off the meter, press the S button. 36 www.i-sens.com www.i-sens.com 37...
  • Page 21: Setting The Alarm Function

    Setting the Alarm Function Four types of alarms can be set in the CareSens N Voice Meter: one Setting the Time Alarms (alarm 1 - 3) post-meal alarm (PP2 alarm) and three time set alarms (alarm 1 - 3). Step 1 The PP2 alarm goes off 2 hours after setting the alarm.
  • Page 22: Caring For Your System

    Step 6 Note: The life span of a CareSens N Voice Meter is 5 years. We Press and hold the S button for 3 seconds to finish and turn the recommend disinfecting both the meter and lancing device at meter off.
  • Page 23: Pre-Cleaning And Disinfection Procedures

    ① Open the cap of the the CareSens N Voice Meter and lancing device. It has been Clorox Germicidal shown to be safe for use with the meter and lancing device. Wipes container and This disinfectant is available commercially in towelette form.
  • Page 24 Never range printed on the test strip vial. See page 19 for how to do immerse the meter or hold it under running water because a control solution test. this will damage the meter. 44 www.i-sens.com www.i-sens.com 45...
  • Page 25: Caution: Storage And Handling

    • The CareSens N Voice Meter should be used only with appeared." appeared. applying the CareSens N Single Test strips. blood or control • Store all meter components in the carrying case to prevent solution sample.
  • Page 26 / Error 8 / Error 9. problem with the meter. Contact Meter failure." meter. Customer Service: 1-800-429-5001. Note: If the error messages persist, contact Customer Service: 1-800-429-5001 (9 am - 9 pm EST, Monday - Saturday). 48 www.i-sens.com www.i-sens.com 49...
  • Page 27: General Troubleshooting

    System has been evaluated in laboratory and clinical tests. • Check whether the test strip is inserted Accuracy: The CareSens N Voice BGM System is calibrated to with the contact bars facing up. Check yield results equivalent to plasma glucose concentrations. The...
  • Page 28 Precision: Precision studies were performed in a laboratory Alternative site test results for glucose concentration ≥ 75 mg/dL using the CareSens N Voice BGMS. Within ± 5% Within ± 10% Within ± 15% Within ± 20% Fingertip 27/48 (56.3%) 46/48 (95.8%)
  • Page 29: Warranty Information

    Warranty Information Manufacturer's Warranty i-SENS, Inc. warrants that the CareSens N Voice Meter shall be free of defects in material and workmanship in normal use for a period of 5 years. The meter must have been subjected to normal use. The warranty does not cover improper handling, tampering, use, or service of the meter.