Precision: The precision studies were performed in a
laboratory using CareSens N POP Blood Glucose Monitoring
Systems.
Within Run Precision
Blood average
Between Run Precision
Control average
This study shows that there could be variation of up to 3.8 %.

Influence Quantities

Packed Cell Volume (Hematocrit)
Packed cell volume evaluation was conducted in various
hematocrit levels. The range of hematocrit levels within the
acceptance criteria is 15–65 %.
2.1 mmol/L
4.6 mmol/L
6.9 mmol/L
11.7 mmol/L
17.1 mmol/L
2.1 mmol/L
6.6 mmol/L
17.4 mmol/L
SD = 0.1 mmol/L
SD = 0.2 mmol/L
CV = 3.0%
CV = 3.2%
CV = 2.6%
SD = 0.1 mmol/L
CV = 3.8 %
CV = 3.3%