• If the meter or test strips do not function properly,
• If your symptoms are inconsistent with the blood glucose
test results and you feel that the meter or test strips are not
working properly,
• If you drop or damage the meter.

Control Solution Testing

Step 1
Insert a test strip into the meter's test strip port,
with the contact bars facing upwards.
Gently push the test strip into the port until the
meter beeps. Be careful not to bend the strip
while pushing it in.
The
symbol will be displayed on the screen.
Step 2
Shake the CareSens Control
Solution bottle well before each
test.
Remove the cap and squeeze the
bottle to discard the first drop.
Then wipe the tip with a clean tissue or cloth. After the
symbol appears on the display, apply the solution to the
narrow edge of the test strip until the meter beeps. Make sure
the confirmation window fills completely.
Note
The meter may switch off if the control solution is not applied
within 2 minutes of the
the meter turns off, remove the strip, reinsert, and start from
step 1.
symbol appearing on the screen. If