i-SENS GM505PAD User Manual

i-SENS GM505PAD User Manual

The accuracy of the

Advertisement

Table of Contents
Information
Important Information: Read This First!
Specifications
CareSens N Blood Glucose Monitoring System
Inserting or Replacing the Batteries
Caring for Your System
CareSens N Blood Glucose Test Strip
CareSens N Blood Glucose Meter
CareSens N Blood Glucose Meter Display
Preparation
Setting Up Your System
Adjusting the Date and Time
Setting the Sound On/OFF
Checking the System
Control Solution Testing
Comparing the Control Solution Test Results
Testing
Using the Lancing Device
Preparing the Lancing Device
Preparing the Meter and Test Strip
Flagging Post-meal Test Results
Applying Blood Sample
Discarding Used Lancets
Alternate Site Testing
HI and Lo Messages
Target Blood Glucose Ranges
Transferring Test Results
Additional Functions
Meter Memory
Viewing Averages Stored in the Memory
Viewing Test Results Stored in the Memory
Setting the Alarm Function
Setting the Post-meal Alarm (PP2 alarm)
Setting the Time Alarms (alarm 1-3)
Maintenance
Understanding Error Messages
General Troubleshooting
Performance Characteristics
Warranty Information
4
6
7
8
9
10
12
13
14
14
17
18
19
21
22
23
25
25
26
28
29
32
33
34
35
35
37
38
38
39
41
43
44
48

Advertisement

Table of Contents
loading

Summary of Contents for i-SENS GM505PAD

  • Page 1 Table of Contents Information Important Information: Read This First! Specifications CareSens N Blood Glucose Monitoring System Inserting or Replacing the Batteries Caring for Your System CareSens N Blood Glucose Test Strip CareSens N Blood Glucose Meter CareSens N Blood Glucose Meter Display Preparation Setting Up Your System Adjusting the Date and Time...
  • Page 2: Important Information: Read This First

    Cautions for safety and optimum product use oxygen therapy. Use by (unopened or opened test strip vial) If you need assistance, please contact your authorised i-SENS Do not discard this product with other household-type waste sales representative or visit www.i-sens.com for more information.
  • Page 3: Specifications

    5-50˚C (41-122˚F) Relative humidity 10-90% • C heck all the components after opening the CareSens N Blood Glucose Monitoring System package. The exact contents are Hematocrit 15-65% listed on the main box. • T he cable for data management software can be ordered separately. Please contact your authorised i-SENS sales representative. 6 www.i-sens.com www.i-sens.com 7...
  • Page 4: Caring For Your System

    Place the cover on the battery compartment. Push down until you hear the tab click into place. Note: Removing the meter batteries will not affect your stored results. However, you may need to reset your meter settings. See pages 14-17. 8 www.i-sens.com www.i-sens.com 9...
  • Page 5: Caresens N Blood Glucose Monitoring System

    Apply blood sample here for testing • D o not bend, cut, or alter test strips in any way. • F or detailed storage and usage information, refer to the CareSens N Test Strip package insert. Caution: • K eep the meter and testing supplies away from young children. • D rying agents in the vial cap may be harmful if inhaled or swallowed and may cause skin or eye irritation. 10 www.i-sens.com www.i-sens.com 11...
  • Page 6: Caresens N Blood Glucose Meter

    It is recommended to check if the display screen on the • T he cable for data management software can be ordered meter matches the illustration above every time the meter separately. Please contact your authorised i-SENS sales turns on. Do not use the meter if the display screen does not representative.
  • Page 7: Setting Up Your System

    The AM/PM is not displayed button to confirm your selection and go in the 24-hour format. to the next step. After selecting the format, press the S button to go to the next step. 14 www.i-sens.com www.i-sens.com 15...
  • Page 8: Setting The Sound On/Off

    OFF. • A t any stage, press the S button for 3 seconds to exit SET mode and turn off the meter. Press and hold to quickly scroll through the numbers. 16 www.i-sens.com www.i-sens.com 17...
  • Page 9: Checking The System

    2 minutes of the symbol appearing on • When using the meter for the first time, the screen. If the meter turns off, remove the strip, reinsert, • Whenever you open a new vial of test strips, and start from step 1. • If the meter or test strips do not function properly, • I f your symptoms are inconsistent with the blood glucose test 18 www.i-sens.com www.i-sens.com 19...
  • Page 10: Comparing The Control Solution Test Results

    CareSens N Test Strip and Meter may not be working properly. Do not use your system and contact your Note: The CareSens Control Solution can be purchased authorised i-SENS sales representative. separately. Please contact your authorised i-SENS sales representative. 20 www.i-sens.com www.i-sens.com 21...
  • Page 11: Using The Lancing Device

    Note: Repeated puncturing at the same sample site may cause pain or skin calluses (thick hard skin). Choose a different site each time you test. 22 www.i-sens.com www.i-sens.com 23...
  • Page 12: Preparing The Meter And Test Strip

    ( ). get an adequate sample size. If you do not want to save the result as a post- meal test, move on to step 9 after step 7. 24 www.i-sens.com www.i-sens.com 25...
  • Page 13: Applying Blood Sample

    3 seconds. shown in the diagram below. Discard used test strips safely in appropriate containers. Good Sample Insufficient Sample 26 www.i-sens.com www.i-sens.com 27...
  • Page 14: Discarding Used Lancets

    Then press the release button while continuing to apply pressure. Keep holding the lancing device against your skin until sufficient (at least 0.5 μL, actual size: ) blood is drawn. Carefully lift the lancing device away from your skin. 28 www.i-sens.com www.i-sens.com 29...
  • Page 15 Note: If the sample drop of blood runs or spreads due to contact with hair or with a line in your palm, do not use that • During two (2) hours after an insulin injection sample. Try puncturing again in a smoother area. 30 www.i-sens.com www.i-sens.com 31...
  • Page 16: Hi And Lo Messages

    American Diabetes Association. “Standards of Medical Care in Diabetes – 2016. ” Diabetes Care. January 2016; 39(1):S15, S100. Note: Please contact your authorised i-SENS sales representative if such messages are displayed even though you do not have hyperglycemia or hypoglycemia.
  • Page 17: Transferring Test Results

    ) results computer using the data cable. from the last 1, 7, 14, 30 and 90 days. For more information, contact your authorised i-SENS sales representative or visit Viewing Averages Stored in the Memory www.i-sens.com. Step 1 Press the or S button to turn the meter on.
  • Page 18: Viewing Test Results Stored In The Memory

    Note: The control solution test results saved with 'check' Note: The control solution test results saved with 'check' are will be displayed with 'check' when you review the stored not included in the averages. test results. 36 www.i-sens.com www.i-sens.com 37...
  • Page 19: Setting The Alarm Function

    Press to set the time. the bell ( ) symbol, and ‘OFF’ will Press to end. appear on the screen. Then the screen will change automatically to the memory check mode without the bell ( ) symbol displayed. symbol disappears 38 www.i-sens.com www.i-sens.com 39...
  • Page 20: Understanding Error Messages

    (5-50˚C/41- 122˚F) and repeat the test after the meter and test strips have reached a temperature within the operating range. 40 www.i-sens.com www.i-sens.com 41...
  • Page 21: General Troubleshooting

    Note: If the problem is not resolved, please contact your sales representative. authorised i-SENS sales representative. Note: If the error messages persist, contact your authorised i-SENS sales representative. 42 www.i-sens.com www.i-sens.com 43...
  • Page 22: Performance Characteristics

    System has been evaluated in laboratory and in clinical tests. using CareSens N BGM Systems. Accuracy: The accuracy of the CareSens N BGM System (Model Within Run Precision GM505PAD, GM505PBD, GM505PCD) was assessed by comparing Blood avg. 2.1 mmol/L SD = 0.1 mmol/L blood glucose results obtained by patients with those Blood avg.
  • Page 23 100% within ± 0.83 mmol/L of the medical laboratory values Hydrocortisone 0.02 mmol/L 1.9% at glucose concentrations below 5.55 mmol/L, and 99.6% within ±15% of the medical laboratory values at glucose Ibuprofen -0.1 mmol/L 2.8% concentrations at or above 5.55 mmol/L. Icodextrin -0.2 mmol/L -0.5% 46 www.i-sens.com www.i-sens.com 47...
  • Page 24: Warranty Information

    Warranty Information Manufacturer’s Warranty MEMO i-SENS, Inc. warrants that the CareSens N Meter shall be free of defects in material and workmanship in normal use for a period of five (5) years. The meter must have been subjected to normal use. The warranty does not cover improper handling, tampering, use, or service of the meter.
  • Page 25 MEMO 50 www.i-sens.com...