Page 2
* The information in this manual is correct at the time of printing. However, i-SENS reserves the right to make any necessary changes at any time without notice as our policy is one of continuous...
Table of Contents Information Important Information: Read This First! Specifications CareSens N POP Blood Glucose Monitoring System Inserting or Replacing the Batteries Caring for Your System CareSens N Blood Glucose Test Strip CareSens N POP Blood Glucose Meter CareSens N POP Blood Glucose Meter Display Preparation Setting Up Your System Adjusting the Date and Time...
Use by (unopened or opened test strip vial) oxygen therapy. Do not discard this product with other household-type waste If you need assistance, please contact your authorised i-SENS Do not reuse Batch code sales representative or visit www.i-sens.com for more information.
POP blood glucose monitoring system package. The exact Hematocrit 15-65% contents are listed on the main box. Storage Conditions • T he cable for data management software can be ordered Meter separately. Please contact your authorised i-SENS sales 0–50˚C (32–122˚F) (with batteries) representative. Test strip 1–30˚C (34–86˚F) 6 www.i-sens.com...
Push it down until you hear local council or authority. If you need assistance, contact your the tab click into place. authorised i-SENS sales representative or visit www.i-sens.com. Note: Removing the meter batteries will not affect your stored results. However, you may need to reset your meter settings.
• D o not bend, cut, or alter test strips in any way. Apply blood sample here for testing • F or detailed storage and usage information, refer to the CareSens N test strip package insert. Caution: • K eep the meter and testing supplies away from young children. • D rying agents in the vial cap may be harmful if inhaled or swallowed and may cause skin or eye irritation. 10 www.i-sens.com www.i-sens.com 11...
Note: unit is set to mmol/L • T he cable for data management software can be ordered separately. Please contact your authorised i-SENS sales Note: It is recommended to check if the display screen on the representative. meter matches the illustration above every time the meter turns • T he unit of measurement is fixed and it cannot be changed...
When the present The AM•PM symbol is not displayed year appears, press the button to in the 24-hour format. After selecting confirm your selection and to go to the the format, press the button to go next step. to the next step. 14 www.i-sens.com www.i-sens.com 15...
If the sound is set to OFF, none of the sound functions will work. After setting the sound, press the button to go to the next step. Note: symbol is displayed only when the sound is set to OFF. 16 www.i-sens.com www.i-sens.com 17...
Note: If the test result is lower than the pre-set hypoglycemia level, the meter will display the following. Caution: Ask your healthcare professional to help you decide what your hypoglycemia level is before setting your level. 18 www.i-sens.com www.i-sens.com 19...
October control solution instead of blood, of 2020, the meter displays 'EP' in the • W hen using the meter for the first time, beginning of November of 2020. 20 www.i-sens.com www.i-sens.com 21...
2 minutes of the symbol appearing Note: The CareSens Control Solution can be purchased on the screen. If the meter turns off, remove the strip, reinsert, separately. Please contact your authorised i-SENS sales and start from Step 1. representative. 22 www.i-sens.com www.i-sens.com 23...
Note: Repeated puncturing at the same sample site may Do not use your system and contact i-SENS sales representative. cause pain or skin calluses (thick hard skin). Choose a different site each time you test.
The skin depth to get blood samples will vary for release button as shown in the diagram. various people at different sample sites. The lancing device's adjustable tip allows the best depth of skin penetration to get an adequate sample size. 26 www.i-sens.com www.i-sens.com 27...
The test result will also be displayed with the post-meal flag ). If you do not want to save the result as a post-meal test, move on to step 9 after step 7. Good Sample Insufficient Sample 28 www.i-sens.com www.i-sens.com 29...
Always dispose of lancets properly. If the test strip is removed after the test result is displayed, the meter will automatically switch off after 3 seconds. Discard used test strips safely in disposable containers. 30 www.i-sens.com www.i-sens.com 31...
Then press the release button while continuing to apply pressure. Keep holding the lancing device against your skin until sufficient (at least 0.5 μL, actual size: ) blood is drawn. Carefully lift the lancing device away from your skin. 32 www.i-sens.com www.i-sens.com 33...
Try puncturing again in a smoother area. Note: Please contact your authorised i-SENS sales representative if such messages are displayed even though you do not have hyperglycemia or hypoglycemia.
Before lunch or dinner data cable. 1 hour after meals For more information, contact your authorised i-SENS sales representative 2 hours after meals or visit at www.i-sens.com. Between 2 a.m. and 4 a.m. Expected Values : The range of a normal fasting* blood glucose level for non-diabetic adults is between 3.9-5.5 mmol/L.
Use the button to scroll back through the averages seen button previously. Average Press and hold the button to turn off the meter. (1, 7, 14, 30, 90 days) button 38 www.i-sens.com www.i-sens.com 39...
To turn off the PP2 alarm, press and hold the button for 3 seconds. 'PP2' and 'OFF' will appear on the screen. Then the button screen will change automatically to the memory recall mode without 'alarm' displayed. button 'alarm' disappears 40 www.i-sens.com www.i-sens.com 41...
The blood or control Repeat the test with a contact your solution sample was new test strip and wait authorised i-SENS sales applied before the until the symbol representative. symbol appeared. appears before applying the blood or...
System accuracy results for glucose concentrations between 2.0 mmol/L and 26.4 mmol/L Note: If the problem is not resolved, please contact your authorised i-SENS sales representative. Within ± 0.83 mmol/L and Within ± 15% 598/600 (99.7%) 44 www.i-sens.com www.i-sens.com 45...
Warranty Information Manufacturer’s Warranty Difference Averages i-SENS, Inc. warrants that the CareSens N POP Meter shall be Interferent Interval1 Interval 2 free of defects in material and workmanship in normal use for (2.8–5.5 mmol/L) (13.9–19.4 mmol/L) a period of five (5) years. The meter must have been subjected to normal use.