Comparing the Control Solution Test Results

The test result of each control solution should be within the
range printed on the label of the test strip vial. Repeat the
control solution test if the test result falls outside of this range.
Out of range results may occur in following situations:
Situations
• W hen the control solution bottle
was not shaken well,
• W hen the meter, test strip, or the
control solution were exposed to
high or low temperatures,
• W hen the first drop of the control
solution was not discarded or the tip
of the bottle was not wiped clean,
• W hen the meter is not functioning
properly.
• W hen the control solution is past the
expiration date printed on the bottle,
• W hen the control solution is past its
discard date (the date the bottle was
opened plus three (3) months),
• W hen the control solution is
contaminated.
If results continue to fall outside the range printed on the test
strip vial, the test strip and meter may not be working properly.
Do not use your system and contact your authorised i-SENS
sales representative.
24 www.i-sens.com
Do This
Repeat the
control solution
test by referring
to the "Notes"
on page 20.
Discard the used
control solution
and repeat the
test using a new
bottle of control
solution.

Using the Lancing Device

You will need a lancing device in order to collect a blood
sample.
You may use the lancing device that is included in the CareSens
N Voice Blood Glucose Monitoring System or any other
medically approved lancing device.
Adjustable Tip
Lancet Holder
Load Confirmation
Lancet
Window
• T he lancing device is for use by a single user only and should
not be shared with anyone.
• U se a soft cloth or tissue to wipe the lancing device.
If necessary, a small amount of alcohol on a soft cloth or
tissue may be used.
Caution:
To avoid infection when drawing a sample, do not
use a lancet more than once, and:
• D o not use a lancet that has been used by others.
• Always use a new sterile lancet.
• Keep the lancing device clean.
Note:
Repeated puncturing at the same sample site may
cause pain or skin calluses (thick hard skin). Choose a
different site each time you test.
Ejection System
Sliding Barrel
Release Button