Fig 1-6 Dynamic Wave Form - Prodigit 34000A Series Operation Manual

High power electronic load
Table of Contents

Advertisement

8 PRODIGIT
1.1.5. Dynamic Waveform Definition
Along with static operation the 34000A Series Electronic Load are built with a Dynamic mode for
operation in Constant Current (CC), Constant Resistance (CR) or Constant Power (CP). This
allows the test engineer to simulate real world pulsing loads or implement a load profile that varies
with time.
A dynamic waveform can be programmed from the front panel of the 34000A Electronic Load.
The user would first set a High and low value of load current using the Level button. The Dynamic
Setting then allows for the rise and fall time between these 2 current values to be adjusted. The
time period that the waveform is high (Thigh) along with the time period that the waveform is low
(Tlow) can also be set.
The dynamic waveform is illustrated below in Fig 1-6.
LOAD
CURRENT
RISE
SLEW RATE
The dynamic waveform can also be set up via the optional computer interface. Dynamic
waveform settings made from the front panel of the load module can also be saved in the
memory of the 34000A series Electronic Load. For the store/recall procedure and the computer
command set please refer to the relevant operating manual for the 34000A series Electronic Load.
Further dynamic waveform definitions are:
The analogue programming input also provides a convenient method of implementing a dynamic
waveform. Please see the section 3.1.26 titled 'Analog Programming Input' for further information.
I
T
HIGH

Fig 1-6 Dynamic Wave form

The period of dynamic waveform is Thigh + Tlow
The dynamic frequency = 1 /( Thigh + Tlow )
The duty cycle = Thigh / ( Thigh + Tlow )
Example1:
343XXA Series, Dynamic up to 50 KHz frequency
Dynamic highest frequency 50 KHz = 0.02ms=20us
Setting THIGH=10 uS, TLOW=10uS, THIGH+TLOW=20uS
CCH-CCL/SR≦10uS
Setting CCH=30A, CCL=10A
(30-10)/2.5A/uS≦10 uS
8 uS≦10 uS ,Compliance with frequency 50KHz
Example2:
Setting THIGH=10 uS, TLOW=10uS, THIGH+TLOW=20uS
CCH-CCL/SR≦10uS
Setting CCH=50A, CCL=0A
(50-0)/2.5A/uS=20uS, 20uS>10uS, It's not compliance the frequency 50 KHz
FALL SLEW RATE
T
LOW
HIGH LOAD
LOAD
LOW LOAD
LEVEL
V