158 PRODIGIT
5.17.3. TYPE 3:
When testing loads on of Constant current (CC) mode or Constant Power (CP), setting
discharge Time, when Load on time achieve setting time, this moment load on auto
change to load off and Display voltage, setting Timer value range 1 to 99999 sec (27H)
(Fig5-15).
Note: Pre-setting the LOAD OFF voltage can protect the battery from discharging when
The preset discharge time has not been reached, and stop the battery when the battery
Voltage is to low to avoid battery damage.
5.17.4. TYPE 4:
Cycle Life test, Battery discharge test use pulse mode, Dynamic mode use count test
And Repeat function, as show Fig 5-18, load on and dynamic on until counter to 0, load
On and dynamic on auto change to off, Remote will shows "OK" and XX.XXX"(V meter),
Cycle setting range 1 to 2000, step setting value 1 to 3 and Repeat setting value 0 to
9999, the setting is by remote operation.
Note: Pre-setting the LOAD OFF voltage can protect the battery from discharging when
The preset discharge time has not been reached, and stop the battery when the battery
Voltage is to low to avoid battery damage.
5.17.5. TYPE 5:
RAMP Mode, Slew rate load on and Repeat function, as Fig 5-19 Show. Setting" STEPn"
n-1~9, CC0, CC1, ΔT1, CC2, ΔT2......CC9, ΔT9, Repeat, Load on mode, Increased or
Decreased current values by every seconds.
ΔCC =(CCn-(CCn-1))/Time, Time:0~6000Sec, STEP: 1~9, Repeat: 0~9999, Load on
Auto change to off and remoter will shows "OK" and XX.XXX"(V meter).
Note: Pre-setting the LOAD OFF voltage can protect the battery from discharging when
The preset discharge time has not been reached, and stop the battery when the battery
Voltage is to low to avoid battery damage.

Fig 5-18 TYPE 4 Battery Discharge Figure