Constant Resistance Mode Application - Prodigit 34000A Series Operation Manual

High power electronic load
Table of Contents

Advertisement

5-5.

Constant Resistance mode application

Operating in Constant Resistance mode is useful for testing both voltage and current
sources. The CR mode is particularly suited for the 'soft start' of power supplies.
This is explained in more detail below.
5.5.1
Power supply power up sequence
In constant current mode the demand at initial 'Load ON' of the preset current value
is almost instantaneous. This might cause the Device under Test (DUT) problems
meeting the relatively high current demand at initial switch on.
For example: A 5V/50A output power supply may not be able to deliver 50A over its
entire start-up range of 0-5 volts. In many cases the power supply's short circuit or
over current protection circuit cause the power supply to shut down. This is because
the power supply is trying to deliver the 50A at a voltage level that is too low.
The answer to this problem is not to use CC mode but to use CR mode instead.
This is because in CR mode the current and voltage ramp up together providing a
'soft start' when compared to standard CC mode.
However please note that with the 34000A series of Electronic Loads allow an
adjustable current ramp can be set. This feature is found within the dynamic settings
as RISE slew rate. Even in static mode the 34000A series load will regulate its
current demand at 'Load ON' in line with the adjusted RISE slew rate. The FALL
slew rate also in the dynamic settings allows the current ramp down to be controlled
at 'Load OFF'.
CURRENT
SOURCE
INPUT
VOLTAGE
Fig 5-6 Constant Resistance mode Application
+
-
ELECTRONIC
LOLAD
CR MODE
V
LOAD RESISTANCE
SETTING
LOAD CURRENT
34000A Series Operation Manual 139
VOLTAGE
SOURCE
ELECTRONIC
LOLAD
CR MODE
-
+
DC
OUTPUT