Grounding Requirements; Environmental Requirements; Repair; Maintenance And Cleaning Methods - Prodigit 34000A Series Operation Manual

High power electronic load
Table of Contents

Advertisement

2-3 Grounding requirements

SHOCK HAZARD
1. It is requested to use the 3Pin plug connector only for 34000A Series mainframe to out of
danger when electric leakage. And the complete and proper grounded is necessary.
2. The 34000A Series high power load is equipped with three conductor cable which plugs in an
appropriate receptacle to ground the instrument's cover.

2-4 Environmental requirements

Indoor use.
Measurement Category I.
Pollution Degree 2.
Relative Humidity 80% Max.
Ambient Temperature 0 to +40°C
Altitude up to 2000m.
The equipment is not for measurements performed for CAT II, III and IV.
Transient Overvoltage on the mains supply can be 2500V.

2-5 Repair

If the instrument is damaged, please attach a tag to the instrument to identify the owner and
indicated the require service or repairing. And inform the Prodigit sales and service office or
representative.

2-6 Maintenance and cleaning methods

The dust on the load can be removed with a brush, and the dust on the vent below the front panel
can be removed with a vacuum cleaner.
You can use a soft or damp cloth to clean the outer casing of the product. The inside of the product
must be cleaned with a low-pressure air gun, or sent to the original factory or agent for cleaning.
* Please clean regularly every year
Use a soft or slightly damp cloth to clean this product.
BEFORE you clean the unit, switch the mains power off and disconnect the input lead.
・ Please do NOT use any organic solvent capable of changing the nature of the plastic
such as benzene or acetone.
・ Please ensure that no liquid is allowed to penetrate this product.
34000A Series Operation Manual 31