Optional Density Calibration - Optex DOS-20 Instruction Manual

Optical dissolved oxygen sensor
Table of Contents

Advertisement

Optional density calibration

It is possible to immerse the sensor in the solution for which the dissolved
oxygen concentration is known and align the display value with this
concentration.
Item
[Display]
Dissolved oxygen
concentration [DO]
Prepare calibration ("Preparation" P.21).
1
Pour the solution with the known
2
concentration into the calibration
container up until the fill line.
Insert the sensor's temperature sensor until it is immersed
3
in the solution.
・ Insert the measurement area so that approximately 1 cm is
Caution
・ Insert the sensor so that no bubbles build up on the
Leave this to stand for at least 5 minutes to allow the
4
temperature to stabilize.
Press
5
The sensor selection display is displayed.
Setting range
2.00 to 50.00 (by 0.01)
left between it and the bottom of the calibration container.
measurement area.
.
Factory default
setting
2.00
27