Cleaning interval (relay)

Set the time interval between cleaning actions.
The settings are to activate the cleaning device using a relay output
("Cleaning output setting" P.20).
Post-
Cleaning
cleaning
activation
standby
Cleaning
Cleaning
device
Measurement
Measurement
value hold
value
Select an item from the following.
For operation, refer to the "Universal Transmitter SC-U1 Operation
Manual", "9 Cleaning" and " ◆ Cleaning Interval (Relay)".
Item
[Display]
Cleaning Interval
(Relay)
[Cleaning Interval-RLY]
, and press
Cleaning interval (relay)
Normal measurement
OFF (No cleaning)
10 to 1440 minutes (by 10 minutes)
.
Cleaning interval (relay)
Post-
Cleaning
cleaning
activation
standby
Cleaning
Measurement
value hold
Setting range
Normal measurement
Factory
default setting
30 minutes