3.8. REMOTE CONTROL
3.8.1. Frequency Control
1.
Select the position "Ext. Volt." on the SWEEP CONTROL.
2.
A DC control voltage (between 0 and 1 0 V) is applied to pin 7 of the FREQUENCY
CONTROL VOLTAGE "In" socket, see Fig.2.6 (see also section 4.8).
3.
The remaining controls on the 2010 are selected as appropriate to the measure-
ment being made.
Note:
If a 8 & T programme is not used, the 8 & T PROGRAM selector should be in
the "Manual" position.
3.8.2. B & T Programmes
Remote control of a B & T programme is possible only when the frequency control vol-
tage itself is provided by an external voltage
.
In this case, the procedure is as follows:
1.
The frequency is controlled according to section 3.8
.
1.
2.
Position the frequency scale pointer in the passive zone.
3.
Choose a suitable programme using the three selectors 8 & T PROGRAM, SELECTIV-
ITY CONTROL, and EFFECTIVE AVERAGING TIME.
4
.
Using the EXT
.
8 & T PROGRAM socket on the rear panel, ground the pin appropri-
ate to the desired range.
35