4. Select a suitable regulation speed using the COMPRESSOR SPEED
switch.
5. Turn the 8.F.O. OUTPUT VOLTAGE knob
slowly clockwise to the
maximum position, checking that the compressor circuit works.
6. Turn the COMPRESSOR VOLTAGE knob anticlockwise until the
desired level is obtained.
4.6. REMOTE CONTROL OF THE 2010
4.6.1. Frequency Control
1. Select the position "Ext. Volt." on the SWEEP CONTROL.
2. The DC control voltage (between 0 and 10 Volts) is applied to pins 6
and
7
of the FREQUENCY CONTROL VOLTAGE "In" socket, see
Fig.4.1.
3. The remaining controls on the 2010 are selected appropriate to the
measurement being made.
Note: If a 8 & T programme is not used, the 8 & T PROGRAM Selector
should be in the "Manual" position, and the frequency scale pointer should
be in the passive zone.
Control Voltage
0-10 V DC
Ground
Fig.4.1. Connection for remote control of frequency by a D.C. Voltage
36