Avoid Buildup; Take Clearance Into Consideration - Endress+Hauser Liquiphant FTL63 Operating Instructions Manual

Point level switch for liquids specifically for the food and life sciences industries
Hide thumbs Also See for Liquiphant FTL63:
Table of Contents

Advertisement

Mounting
12
High viscosity
NOTICE
Highly viscous liquids may cause switching delays.
Make sure that the liquid can run off the tuning fork easily.
Deburr the socket surface.
High viscosity, e.g. viscous oils: ≤ 10 000 mPa⋅s
The tuning fork must be located outside the installation socket!
 6
Installation example for a highly viscous liquid. Unit of measurement mm (in)
5.1.3

Avoid buildup

• Use short installation sockets to ensure that the tuning fork projects freely into the
vessel
• Leave sufficient distance between the buildup expected on the tank wall and the tuning
fork
 7
Installation examples for a highly viscous process medium
5.1.4

Take clearance into consideration

Allow sufficient space outside the tank for mounting, connection and settings involving
the electronic insert.
> 40 (1.57)
Liquiphant FTL63
A0037348
A0033239
Endress+Hauser